Recording Set-up
Nicolet Portable Clinical System
“Cheat sheet” for setting up the clinical system for a grid patient (helpful when testing triggers with portable clinical system machine): Setting Up Second Clinical.docx
Portable Nicolet Clinical System Login Information:\
Username: portable##
Password: nicolet01
- Grab a portable machine from EEG monitoring room (make sure it’s not a sleep study or mapping computer)
- Ensure you have a CLTM amplifier (and not V44) and at least one 64-electrode jackboxs
- Bring equipment into our office
- Turn on the portable and ensure a working Ethernet cable is plugged in
- Log in pw for every portable: nicolet01
- Open up “Nic-Vue”
- Click “New”
- Patient ID: Pt initials and “Clin2” ex: “CB_clin2”
- Insert patient last name and first name
- Insert tech name (your initials or whatever)
- Click “NicoletOne”
- Select protocol (NYU64 or NYU128, whichever is appropriate)
- Click “verify” – if unable to verify, make sure you have the appropriate amplifer and jackbox and that the connection is solid. Also check Ethernet connection
- Once the recording page opens up, click “settings”
- Click “protocol” at bottom of page
- Highlight the appropriate protocol (either NYU C64 or NYU C128, etc)
- Duplicate that protocol, and title it the patient’s name (last name, first name) and “clin2” (ex: “Boomhaur_Callah_Clin2”)
- Save and apply
- Click “grid/strip”
- Click the edit button
Use the “clin 2 montage” for the following:
- Use the drop down menu to choose the number of contacts (ex: 10 contact LTM strip, stainless steel)
- Click the “ADD” button and name it appropriately (for example, “LTM” or “PO” or GA” etc
- Repeat for all grids/strips
- If you need to turn off any contacts, start of the top
- Click the + button to expand the strip/grid
- Uncheck/check the contacts as necessary
- When finished, click the – to unexpand
- Expand the next one down to turn off/on any contacts until completed
- Click “OK”
- Click “save” and name it the name of the patient and “clin 2” (ex: “Boomhaur_Callah_Clin2”)
- Click on “sensors”
- Duplicate the 10-20 group and name patient name +clin2 (ex: “Boomhaur_Callah_Clin2”)
- and if the DC channels are not already there, insert them
- Name – insert for DC1 through DC 12
- type for DC1-8 is “pulse”
- type for DC9-12 is “EXT DC”
- change the color to teal
- Save
- Click on “Amplifer”, make sure all of your electrodes are turned on, that the sensors are correct, and all Sampling Rates are set to “common”
- Select SEEG DC1 through SEEG DC12
- Select and type in the appropriate sensor for each input
- Example”: “SEEG DC1” input will be sensor “DC1”
- Sampling rate for all DC channels are common
- Save and Apply
- Click “Montage”
- On the left, if correct montagbe available select it (and name the pt name and Clin2), if not, click the “new” icon, and label it the patient name + clin2 (ex: “Boomhaur_Callah_Clin2”)
- Make sure it’s using correct reference (example: Active should be “LTM05”, while Reference will be “LTM06”, then the next will be “LTM07” and reference will be “LTM08”
- For electrodes, you want to lick “audio” and not “special”
- Insert DC channels 1-12 under “active”
- For DC channels:
- Reference: should be “Ref” and not “bipolar” (if you do this once you’re good. You will notice, frustratingly, that once you exit out of the settings it reverts back to “bipolar”. Don’t worry about it.
- Color: Teal
- Label: the DC channel (either 1-12, as appropriate)
- DC1-8: click “audio”and not “special”
- DC9-12: click “special” and not “audio”
- Save and apply
- Click “sets” at the bottom of the page
- Click the “New” icon
- Give the set the patient’s name
- Move your montage over from “mantages” to “montages in set”
- Click save
- Click on “Detections” at the bottom of the page
- Highlight LTM
- Verify Spike and Seizure are turned on
- Click on “Misc” at the bottom of the page
- Verify Impedance Test is unchecked
- Click on Protocols” at the bottom of the page
- Verify the montage set is in your patient’s name, the montage is in your patient’s name and detection is LTM
- Click save and apply
- Exit out of Settings
- Select the electrodes (not DC channels), and on the right, change sensitivity to “70.0”
- Click notch at the top of the page
- Select the DC channels, right click
- Selected “selected traces”, then “display type” and select “trace” (rather than numerical)
- Select DC channel 9-12 separately and select “Adjust selected” and change sensitivity to 500 and the deflection to 1000